Recall Z-1740-2025 ยท LLZ System, Image Processing, Radiological

Philips Medical Systems Nederland B.V.

IntelliSpace Cardiovascular software versions 6.x, 7.x, and 8.x; Model Number: 830089.

DateMay 8, 2025 (initiated Apr 7, 2025)
StatusOpen, Classified
ReasonAn issue was identified with the software when using the Echo Module of ISCV, the issue may result in missing or incomplete information in the report.
Root causeSoftware in the Use Environment
ActionOn April 7, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. 1. Actions that should be taken by the customer / user in order to prevent risks for patients or users " Verify that text added into the report conclusion matches the data in the work area interpret sheet (Refer to Attachment A Screenshot 2) " You may continue to use your system(s) in accordance with the intended use and by following the recommendation above. " Circulate this notice to all users of this device so that they are aware of the potential issue. " Please retain this letter with your system(s) until a s
Quantity996 units (267 US, 729 OUS)
DistributionDomestic: Nationwide Distribution International: Australia, Austria, Belgium, Canada, Denmark, Estonia, Finland, France, Germany, Greece, Hong Kong, Indonesia, Ireland, Italy, Japan, Luxembourg, Mexico, Netherlands, Norway, Poland, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Thailand, U
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K153022Philips IntelliSpace Cardiovascular