ClearanceBrief
Recall Z-1438-2025 ยท OEZ Cardiovascular Procedure Kit

MEDLINE INDUSTRIES, LP - Northfield

Medline procedure kits labeled as: 1) AV FISTULA PACK, Pack Number DYNJ80088; 2) AV FISTULA PACK-LF, Pack Number DYNJ0131782J; 3) CABG, Pack Number DYNJ906462B; 4) CABG ACCESSORY PACK-LF, Pack Number DYNJ0261346U; 5) CARDIAC CATH PACEMAKER-LF, Pack Number DYNJ42997C; 6) CARDIAC CATH PACK, Pack Number DYNJ61256A; 7) CARDIAC CATH PACK-LF, Pack Number DYNJ61527; 8) CARDIAC CATH PACK-LF, Pack

DateMar 25, 2025 (initiated Feb 11, 2025)
StatusOpen, Classified
ReasonMedical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.
Root causeUnder Investigation by firm
ActionMedline Industries LP issued a recall notice to its consignees on 02/11/2025 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other prob
Quantity3860 units
DistributionWorldwide distribution - US Nationwide and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.