Recall Z-1462-2025 ยท FTN Plastic Surgery And Accessories Kit
MEDLINE INDUSTRIES, LP - Northfield
Medline procedure kits labeled as: 1) BREAST PACK-LF, Pack Number DYNJ0956628O; 2) BREAST SURGERY FOAKS, Pack Number DYNJ63577C; 3) CHEST BREAST -LF, Pack Number DYNJCEN04C; 4) DIEP PACK, Pack Number DYNJ60669G; 5) FACE LIFT PACK, Pack Number DYNJ64670B; 6) FREE FLAP A PACK, Pack Number DYNJ40647G; 7) IPMC BREAST PACK-LF, Pack Number DYNJ0665560D; 8) OCULAR PLASTICS PACK, Pack Number DYNJ
| Date | Mar 26, 2025 (initiated Feb 11, 2025) |
| Status | Open, Classified |
| Reason | Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health. |
| Root cause | Under Investigation by firm |
| Action | Medline Industries LP issued a recall notice to its consignees on 02/11/2025 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other prob |
| Quantity | 1228 units |
| Distribution | Worldwide distribution - US Nationwide and the country of Canada. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.