Recall Z-1461-2025 ยท OGR Ear, Nose, And Throat Surgical Tray
MEDLINE INDUSTRIES, LP - Northfield
Medline procedure kits labeled as: 1) BILLINGS CLINIC SC HEAD & NECK, Pack Number DYNJ33656C; 2) BLEPH/RHINO PACK, Pack Number DYNJ64669A; 3) DB ENT CSTM PACK-LF, Pack Number DYNJ33585C; 4) DENTAL PACK, Pack Number DYNJ54064; 5) EAR PACK, Pack Number DYNJ68678C; 6) ENT HEAD NECK PACK, Pack Number DYNJ49023D; 7) ENT PACK, Pack Number DYNJ42564B; 8) ENT PACK, Pack Number DYNJ50572; 9) ENT P
| Date | Mar 26, 2025 (initiated Feb 11, 2025) |
| Status | Open, Classified |
| Reason | Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health. |
| Root cause | Under Investigation by firm |
| Action | Medline Industries LP issued a recall notice to its consignees on 02/11/2025 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other prob |
| Quantity | 2161 kits |
| Distribution | Worldwide distribution - US Nationwide and the country of Canada. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.