Recall Z-1303-2025 ยท JAK System, X-Ray, Tomography, Computed

GE Medical Systems, LLC

GE Healthcare Revolution CT ES, System, X-ray, Tomography, Computed

DateMar 7, 2025 (initiated Feb 18, 2025)
StatusOpen, Classified
ReasonGE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
Root causeProcess change control
ActionGE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/18/2025 via letter delivered using a traceable means. The notice explained the issue, safety issue, and requested the following actions be taken: "You can continue to use your system. Please monitor the floor near the base of the gantry for any accumulation of blue fluid. If blue fluid is observed, wear nitrile or rubber gloves, clean up the fluid by absorbing it with a noncombustible absorbent material, and dispose of the absorbent material in a suitable container for disposal. Please contact your local
Quantity21 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163213Revolution CT
Ge Medical Systems, LLC, Dec 16, 2016
K191777Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT with Apex edition
Ge Medical Systems, LLC, Jul 26, 2019
K213715Revolution Apex, Revolution Apex Elite, Revolution Apex Plus, Revolution Apex Select,Revolution CT; Revolution CT ES; Re
Ge Medical Systems, LLC, Dec 17, 2021