Recall Z-1431-2025 ยท OEZ Cardiovascular Procedure Kit

MEDLINE INDUSTRIES, LP - Northfield

namic convenience kits labeled as: 1) MTS,4 VALVE,LFT,HRT,KIT-AHS-A*,NW, REF 60010468; 2) MTS,RIGHT HEART,KIT,W/TRANS,ST. *, REF 60010511; 3) MTS,LT,HT,KIT-BRYAN,LGH,M.C.,EAST, REF 60020323; 4) MTO,RIGHT,KIT-BRONSON,M* HOSP PG, REF 60020676; 5) MTS,LEFT HEART,KIT-C*,MEM HLTHCARE, REF 60035977; 6) MTS,LEFT HEART,KIT,J*,M*,C*, REF 60040273; 7) MTS,LEFT HEART,KIT,C*,C*, REF 60100313; 8) MTS,

DateMar 21, 2025 (initiated Feb 12, 2025)
StatusOpen, Classified
ReasonThe female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near the fluid pathway.
Root causeProcess change control
ActionMedline Industries issued two notices, delivered using email and US Mail: 1) a MEDICAL DEVICE RECALL notice for the affected Pressure Monitoring Lines and 2) a MEDICAL DEVICE RECALL notice for the kits containing the affected product; both notices explained the issue and the hazard. The MEDICAL DEVICE RECALL notice for the affected Pressure Monitoring Lines requested that the consignee destroy any affected inventory and reply to receive credit. The MEDICAL DEVICE RECALL notice for the kits containing the affected product requested the consignee respond to the notice and upon response, Medlin
Quantity1669 units
DistributionWorldwide distribution - US Nationwide and the countries of The Netherlands, New Zealand, Slovakia