ClearanceBrief
Recall Z-1430-2025 ยท OEQ Angiography/Angioplasty Kit

MEDLINE INDUSTRIES, LP - Northfield

namic convenience kits labeled as: MTS,COLOR,PML,KIT,-,FRENCH HOSPITAL PG; medical convenience kit, REF 600605710

DateMar 21, 2025 (initiated Feb 12, 2025)
StatusOpen, Classified
ReasonThe female luer fittings on the Pressure Monitoring Lines (PMLs) were manufactured with excess material near the fluid pathway.
Root causeProcess change control
ActionMedline Industries issued two notices, delivered using email and US Mail: 1) a MEDICAL DEVICE RECALL notice for the affected Pressure Monitoring Lines and 2) a MEDICAL DEVICE RECALL notice for the kits containing the affected product; both notices explained the issue and the hazard. The MEDICAL DEVICE RECALL notice for the affected Pressure Monitoring Lines requested that the consignee destroy any affected inventory and reply to receive credit. The MEDICAL DEVICE RECALL notice for the kits containing the affected product requested the consignee respond to the notice and upon response, Medlin
Quantity100 units
DistributionWorldwide distribution - US Nationwide and the countries of The Netherlands, New Zealand, Slovakia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.