Recall Z-1407-2025 ยท LLZ System, Image Processing, Radiological

FUJIFILM Healthcare Americas Corporation

Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200.

DateMar 20, 2025 (initiated Feb 11, 2025)
StatusOpen, Classified
ReasonThe incorrect computed patient age is showing in VX for patients less than 3 months old.
Root causeSoftware change control
ActionOn February 11, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. ACTIONS TO BE TAKEN BY THE COMPANY Fujifilm will issue corrections for all impacted 7.4.x versions. The following patch releases will be made available: - 7.4.020 for 7.4.0x versions (immediately available) - 7.4.120 for 7.4.1x versions (available February 28, 2025) - 7.4.210 for 7.4.2x versions (available February 28,2025) ACTIONS THAT CAN BE TAKEN BY THE END-USER IMMEDIATELY Regarding this issue, Synapse PACS VX can safely be used by the end-user as long as; 1] the end-user is aware of the issue and, 2] takes
Quantity104 units (69 US, 35 OUS)
DistributionDomestic: AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, KY, LA, MA, MN, MO, MS, NC, NJ, NY, OH, OK, OR, TN, TX, VT, WA, WI, WV, WY. International: Australia, Belgium, Brazil, France, Germany, Portugal, UK
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160108Synapse PACS
Fujifilm Medical Systems U.S.A, Inc., Feb 3, 2016
K190232Synapse PACS
Fujifilm Corporation, Mar 26, 2019