ClearanceBrief
Recall Z-1398-2025 ยท LLZ System, Image Processing, Radiological

Merge Healthcare, Inc.

VERICIS, Merge Cardio, Model/Catalog Number: Version 9.x, Software Version: 9.0.6 and 9.0.8; VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.

DateMar 20, 2025 (initiated Feb 19, 2025)
StatusOpen, Classified
ReasonWhen following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
Root causeSoftware design
ActionMerge Healthcare issued an Urgent Field Safety Notice to it's consignees on 02/19/2025 via email. The notice explained the issue, risk to patients, provided a workaround to immediately address the issue, and recommended clinicians check measurements in the final report to ensure that they accurately reflect the patient's condition pending software update. Required Actions: Merge Healthcare Support will contact all customers to: i) Confirm that you have read and understood this Urgent Field Safety Notice. ii) Schedule an update to implement the configuration change on your system. For mo
Quantity2
DistributionUS
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051649VERICIS CARDIOVASCULAR IMAGE AND INFORMATION SYSTEM, MODEL 4.0
Camtronics Medical Systems, Aug 2, 2005