Recall Z-1492-2025 ยท DRY Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corporation

Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754

DateMar 31, 2025 (initiated Mar 3, 2025)
StatusOpen, Classified
ReasonDuring some clinical cases, it has been observed that the SO2 value may periodically appear as dashes on the Touchscreen Display indicating unavailable data. In these cases, a yellow low-limit alarm will also be triggered to the user, due to the dashed-out value.
Root causeSoftware design
ActionTerumo Cardiovascular issued an URGENT MEDICAL DEVICE REMOVAL notice to its consignees on 03/03/2025 via email. The notice explained the reason for the removal, potential hazard, and requested the following: 1. Review this Medical Device Correction and assure that all users have received notice of this issue. 2. Confirm receipt of this communication by completing and returning the attached Customer Response Form as indicated on the form. 3. Terumo CVS will contact users to discuss removal and correction activities. Note: Terumo CVS recommends that users continue using CDI OneView System
Quantity37 units
DistributionUS: IL, TX, NY, MI, FL, NV, and Australia, New Zealand
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K234065CDI OneView Monitoring System
Terumo Cardiovascular Systems Corporation, Apr 25, 2024