Recall Z-1385-2025 ยท CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase
Oridion Medical 1987 Ltd.
Microstream Instructions for Use and Part Number used with - Product Description, REF: Microstream Advance Intubated Infant, PT00156250 - Microstream Advance Neonatal-Infant Intubated CO2 Filter Line 2m/6.5ft Extended Duration, MVIIH; Microstream Advance Neonatal-Infant Intubated CO2 Filter Line 2m/6.5ft High Humidity, MVIIHH; Microstream Advance Neonatal-Infant Intubated CO2 Filter Line 4m/
| Date | Mar 17, 2025 (initiated Feb 17, 2025) |
| Status | Open, Classified |
| Reason | Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation. |
| Root cause | Labeling design |
| Action | On 2/17/2025, correction notices were mailed to Risk Manager, Director of Respiratory Care, Director of Anesthesiology who were informed of the following: In September 2025, the firm will issue an update to the Instructions for Use (IFUs) for the The following IFU's: 01158, PT00156355, PT00156254, PT00156250 stating: Adverse events associated with attaching and detaching the airway adapter from the breathing circuit are listed in descending order of severity: unintended extubation, respiratory failure, hypoxia, low oxygen saturation, aspiration/inhalation and delay to treatment. Any serious i |
| Quantity | 704423 |
| Distribution | Worldwide - US Nationwide distribution including in the states of TX, OH, FL, CO, IL, CA, MD, PA, VA, WI, IN, MI, NY, GA, OR, SC, NC, PR, AZ, KY, AR, TN, MO, DE, MA, MS, UT, DC, NE, IA, WV, NJ, KS, LA, WA, HI, MN, AL, ME, RI, OK, MT, ID, CT, NH, AK, NV, SD, WY, ND, NM and the countries of Algeria, A |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K181467 | Microstream Luer Adult-Pediatric Intubated Sampling Line, Microstream Advance Adult-Pediatric Intubated Filter Line Oridion Medical 1987 , Ltd., Feb 8, 2019 |