ClearanceBrief
Recall Z-1384-2025 ยท CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase

Oridion Medical 1987 Ltd.

Microstream Instructions for Use and Part Number used with - Product Description, REF: Microstream Advance Intubated Adult-Pediatric, PT00156254 - Microstream Advance Adult-Pediatric Intubated CO2 Filter Line 2m/6.5ft Short-term use: Procedural/Emergency, MVAI; Microstream Advance Adult-Pediatric Intubated CO2 Filter Line 2m/6.5ft Short-term use: Procedural/Emergency, MVAI100U; Microstream

DateMar 17, 2025 (initiated Feb 17, 2025)
StatusOpen, Classified
ReasonNeonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
Root causeLabeling design
ActionOn 2/17/2025, correction notices were mailed to Risk Manager, Director of Respiratory Care, Director of Anesthesiology who were informed of the following: In September 2025, the firm will issue an update to the Instructions for Use (IFUs) for the The following IFU's: 01158, PT00156355, PT00156254, PT00156250 stating: Adverse events associated with attaching and detaching the airway adapter from the breathing circuit are listed in descending order of severity: unintended extubation, respiratory failure, hypoxia, low oxygen saturation, aspiration/inhalation and delay to treatment. Any serious i
Quantity8435772
DistributionWorldwide - US Nationwide distribution including in the states of TX, OH, FL, CO, IL, CA, MD, PA, VA, WI, IN, MI, NY, GA, OR, SC, NC, PR, AZ, KY, AR, TN, MO, DE, MA, MS, UT, DC, NE, IA, WV, NJ, KS, LA, WA, HI, MN, AL, ME, RI, OK, MT, ID, CT, NH, AK, NV, SD, WY, ND, NM and the countries of Algeria, A
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181467Microstream Luer Adult-Pediatric Intubated Sampling Line, Microstream Advance Adult-Pediatric Intubated Filter Line
Oridion Medical 1987 , Ltd., Feb 8, 2019