Recall Z-1468-2025 ยท IOQ Bed, Flotation Therapy, Powered

Baxter Healthcare Corporation

Progressa Bed Surfaces, intended to be used to treat or prevent pulmonary or other complications associated with immobility, Product Codes P7520A19, P7520A20S, P7520A21, P7520A22S, P7520A23, P7520A24S, P7520A3, P7520A31, P7520A37, P7520A39, P7520A45, and P7520A4S.

DateMar 26, 2025 (initiated Feb 25, 2025)
StatusOpen, Classified
ReasonThe air bladders inside the mattress may move out of position when the head of the bed is elevated, causing a dip in the mattress.
Root causeNonconforming Material/Component
ActionAn URGENT MEDICAL DEVICE CORRECTION notification letter dated 2/25/25 was sent to customers. Actions to be Taken by Customers 1. Please locate all affected units within your facility by reviewing the product label at the head end of the Progressa bed surface with the affected product details provided above. Baxter recommends identifying patients at a higher risk for pressure injuries and placing these patients on alternate beds, if available. 2. A Baxter representative will contact your facility to schedule the correction of your impacted surfaces, or you may contact the Baxter Technical Su
Quantity2,126 units
DistributionUS Nationwide. Global Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122473TOTALCARE SPOT TOTALCARE BARIATRIC TOTALCARE DUO TOATALCARE BED SYSTEM TOTALCARE DUO2
Hill-Rom, Jun 3, 2013