ClearanceBrief
Recall Z-1267-2025 ยท LLZ System, Image Processing, Radiological

MIM Software Inc

MIM software; System, Image Processing, Radiological

DateMar 4, 2025 (initiated Feb 19, 2025)
StatusOpen, Classified
ReasonIn situations where two images with differing Fields of View (FOV) complete an image fusion, an incorrect, elevated Maximum Standardized Uptake Value (SUV) within MIM Software versions 7.2.0 through 7.2.6.could result.
Root causeSoftware design
ActionGE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/19/2025 disseminated using a traceable means. The notice explained the issue, potential risk to the patient, and provided the following actions to be taken to use the device: "You can continue to use your device if you are not using the SUV tool or if you access a non-impacted version of MIM when using the SUV tool. If using a non-default workflow and fusing a PET or SPECT image with another image with a different FOV, do not incorporate the Maximum SUV value into assessment of disease status, progressio
Quantity352 units
DistributionWorldwide distribution - United States Nationwide and the countries of Australia, Belgium, Bolivia, Canada, China, Croatia, Czech Republic, Denmark, France, Germany, Greece, Israel, Italy, Japan, Mexico, Netherlands, Poland, Russia, Spain, Sweden, United Kingdom, Finland, Hong Kong.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071964MIM 4.1 (SEASTAR)
Mimvista Corp., Sep 26, 2007