ClearanceBrief
Recall Z-1418-2025 ยท JEA Table, Surgical With Orthopedic Accessories, Ac-Powered

Baxter Healthcare Corporation

Baxter TruSystem 7500 Hybrid (SC), Product Code 1854086

DateMar 21, 2025 (initiated Feb 25, 2025)
StatusOpen, Classified
ReasonThere is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
Root causeSoftware design
ActionBaxter issued an UREGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/25/2025 by letter sent via traceable means. The notice explained the issue, hazard involved, and requested the following actions be taken: "Actions to be Taken by Customers 1. Baxter will contact you to install a software upgrade in the impacted surgical table systems. 2. If you received this communication directly from Baxter, acknowledge receipt by following the instructions on the enclosed reply form instruction sheet, even if you have no remaining inventory. Acknowledging receipt of this notification will
Quantity8 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.