Recall Z-1294-2025 ยท GAR Suture, Nonabsorbable, Synthetic, Polyamide

TAS Medical Inc

TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP

DateMar 6, 2025 (initiated Jan 27, 2025)
StatusOpen, Classified
ReasonTissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation.
Root causeDevice Design
ActionOn 1/27/2025, correction notices were emailed to customers who were asked to do the following: 1) Firm has developed and validated an interim method for conditioning the product in the surgical suite just prior to implantation to rehydrate the zip-tie straps per these instructions. To reverse the effects of dehydration, when preparing for surgery, submerge the TAS zip-tie straps in 40-degreesC (104-degrees F) sterile saline solution for 60 minutes prior to implantation. The orange leader material is not subject to dehydration and therefore is not required to be soaked. 2) Quarantine affec
Quantity114
DistributionUS Nationwide distribution in the states of LA, OK, TN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K220980Tissue Approximation System (TAS)
Tas Medical, Inc., Jul 5, 2023