Recall Z-1387-2025 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter, Inc.

DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, with Component PCB, TSI/RSI firmware C29278

DateMar 17, 2025 (initiated Feb 6, 2025)
StatusOpen, Classified
ReasonBeckman Coulter identified that an issue for DxI 9000 tip stocker interface (TSI) and reagent storage interface (RSI) boards with the firmware version 1.1.0.0 can cause sub-system communication failures. This failure could result in a delay in reporting patient results.
Root causeDevice Design
ActionAn URGENT MEDICAL DEVICE RECALL notification letter was sent to customers beginning on 2/6/25. Beckman Coulter has determined that a DxI 9000 Access Immunoassay Analyzer containing tip stocker interface (TSI) and reagent storage interface (RSI) boards with the firmware version 1.1.0.0 can cause sub-system communication failures. If the analyzer experiences the issue described above, it can lead to the analyzer entering the red, or Stopped, system status due to Reagent Storage and Tip Stocker "axis is out of tolerance" errors. Any in process tests that require addition of reagent at the tim
Quantity80 units
DistributionWorldwide - US Nationwide distribution in the states of FL, TX, WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K221225Access TSH (3rd IS) Assay, DxI 9000 Access Immunoassay Analyzer
Beckman Coulter, Inc., Nov 10, 2022