Recall Z-1289-2025 ยท BTR Tube, Tracheal (W/Wo Connector)

Smiths Medical ASD, Inc.

Portex Endotracheal Tube, Item Codes 100/105/025 100/105/030 100/105/035 100/111/020 100/111/025 100/111/030 100/111/035 100/111/035 100/126/025 100/126/030 100/126/035 100/127/025 100/127/030 100/127/035 100/141/025 100/141/030 100/141/035

DateMar 12, 2025 (initiated Feb 13, 2025)
StatusOpen, Classified
ReasonAffected devices have a smaller diameter than expected and may potentially result in inadequate ventilation to the patient. Patient may experience hypoxia, underdose, or cardiopulmonary collapse.
Root causeProcess change control
ActionOn February 13, 2025, the firm began to notify affected consignees via Urgent Medical Device Correction letters. Customers were informed that affected sizes of endotracheal tube products may have a smaller diameter than expected. Customers were instructed to check all inventory locations for affected products and discontinue use. Affected product should be destroyed following your institution's process for destruction. Product should be quarantined until disposal. All potential users of the device should be informed. If you have distributed affected product, please immediately forward the re
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K790312ENDOTRACHEAL TUBE W/HOLDER & CONNECTOR
Respiratory Support Products, Inc., Feb 21, 1979