Recall Z-1291-2025 ยท GAZ Tubing, Noninvasive

DeRoyal Industries Inc

DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C7100NS Tubing, Suction, Connector Indicated use - Fluid Movement

DateMar 5, 2025 (initiated Jan 31, 2025)
StatusOpen, Classified
ReasonDue to the inner diameter of the suction tubing connector being smaller than the suction canister lid port connection. This size difference, combined with the durometer of the plastic material, makes the assembly to a suction canister lid port difficult. This could result in inadequate suction and cause a delay in patient care.
Root causeUnder Investigation by firm
ActionOn 01/31/2025, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter via email to customers informing them that bulk roll, non-sterile suction connector tubing products and lots which could possibly be defective. The inner diameter of the suction tubing connector, is smaller than the suction canister lid port connection. This size difference, combined with the durometer of the plastic material, makes the assembly to a suction canister lid port difficult. This could result in inadequate suction and cause a delay in patient care. Customers are instructed to: 1) Identify affected product numbe
Quantity714 cases
DistributionU.S.: IL, MN, NC, PA, and VA O.U.S.: N/A
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.