Recall Z-1319-2025 ยท MKJ Automated External Defibrillators (Non-Wearable)
ZOLL Medical Corporation
Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-01C, (2) G5A-02C, (3) G5A-05C, (4) G5A-08C, (5) G5A-10C, (6) G5A-13C, (7) G5A-15C, (8) G5A-22C, (9) G5A-41C, (10) G5A-85C, (11) G5A-90C (12) G5A-80-L; Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, su
| Date | Mar 12, 2025 (initiated Feb 13, 2025) |
| Status | Open, Classified |
| Reason | The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity). |
| Root cause | Unknown/Undetermined by firm |
| Action | On February 13, 2025, URGENT: MEDICAL DEVICE CORRECTION letters were sent to customers. REQUIRED ACTIONS Customers who have AEDs should routinely: (1) Monitor device for self-test results. (2) For customers who experience a self-test failure, take action to resolve any alerts as soon as possible. Customers who have at-risk devices should immediately: (1) Communicate this notification to G5 AED users. (2) Ensure users have a program in place that periodically monitors the status of these devices, in accordance with G5 User s Guide. (3) EC 501 is likely due to exposure to humidity outside the de |
| Quantity | 960 units |
| Distribution | Domestic: Nationwide Distribution International: AE, AI, AM, AR, AT, AU, BD, BE, BG, BH, BM, BN, BS, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DO, EG, ES, FI, FR, GB, GH, GU, HK, HR, HU, ID, IE, IL, IN, IS, IT, JE, JM, JP, KR, KW, KZ, LB, LT, MA, MC, MD, ME, MQ, MS, MT, MX, MY, NC, NG, NL, NO, NZ, OM, PA |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.