Recall Z-1405-2025 ยท CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase

Smiths Medical ASD, Inc.

CAPNOCHECK BCI Airway Adapter; Model Number BSR; Product Code WW8214

DateMar 20, 2025 (initiated Feb 14, 2025)
StatusOpen, Classified
ReasonAirway adapter does not meet specification and may break off when attaching a valve or tube. Device failure may result in inaccurate readings, a break in airway circuit, and loss of ventilation. A break in the airway circuit could result in inability to ventilate, hypoxia, cardiopulmonary collapse, or death.
Root causeProcess control
ActionOn February 14, 2025, customers were notified via Urgent Medical Device Correction letters. Consignees were instructed to locate and quarantine all affected product within their institution and destroy them. The recalling firm will contact customers who have returned completed response forms identifying affected products within their possession to discuss credit.
Quantity175,994 US; 45,223 OUS
DistributionUS Nationwide distribution. International distribution pending.
FDA recordOpen on fda.gov

510(k) clearances named in this recall