Recall Z-1506-2025 ยท LKK Pump, Infusion, Implanted, Programmable

Medtronic Neuromodulation

A820 myPTM Software Application associated with Medtronic SynchroMed Pump and Infusion System.

DateApr 1, 2025 (initiated Jan 9, 2025)
StatusOpen, Classified
ReasonProduct complaints were received describing the A820 myPTM app taking longer than expected for patients to interact with their implantable pump.
Root causeUnder Investigation by firm
ActionMedtronic issued an URGENT: MEDICAL DEVICE correction notice to its consignees on 02/28/2025 via email or US mail. The notice explained the issue and provided recommendations to temporarily mitigate the problem, and requested the following: Actions " Share this notice with all those who need to be aware of this issue within your organization and maintain a copy of this notice in your records. " Please provide a copy of the attached patient letter to your affected patients. " Please complete and return the Customer Confirmation Form enclosed with this letter, acknowledging that you have rece
Quantity4365 units
DistributionWorldwide distribution - US Nationwide and the countries of Austria, Belgium, Canary Islands, Czech Republic, France, Germany, Greece, Ireland, Italy, Netherlands, Norway, Portugal, Puerto Rico, Spain, Sweden, Switzerland, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.