The MSOT Acuity Echo is a Class 4 medical laser device. The MSOT Acuity is a mobile assembly of electrically-powered components intended for imaging and analysis of soft-tissue and vasculature using combined optoacoustic (photoacoustic) and ultrasound (US) imaging.

DateFeb 7, 2025 (initiated Jan 8, 2025)
StatusOpen, Classified
ReasonThe MSOT Acuity Echo does not include an adequate instruction for safe operation of the device. The previous version of the user manuals contained a wrong value for the Nominal Ocular Hazard Distance (NOHD).
Root causeRadiation Control for Health and Safety Act
ActionThe firm sent an "Urgent Field Safety Notice" letter dated 12/19/2024 to its consignees. FSN 3746 has been distributed to all customers to correct a wrong value in the user manual for the NOHD. The letter described the product, problem and actions to be taken. The consignees were instructed to do the following: Take note of amendment/reinforcement of Instructions For Use (IFU) 1. Confirm the receipt of this FSN 2. Forward this information and the updated IFU to the person responsible for laser safety at your location (such as Laser Safety Officer) 3. The person responsible for laser safet
Quantity23 units (6 units US)
DistributionWorldwide distribution: US (Nationwide) and OUS (foreign) to countries of: Austria, Germany, Italy, Netherlands, Portugal, Switzerland, and United Kingdom.