ClearanceBrief
Recall Z-1225-2025 ยท HRS Plate, Fixation, Bone

Tyber Medical

A.L.P.S. mvX- MEDIAL TIBIA PLATE 16H LT-Medial Tibia Plate, 16-Hole, Left Intended for Fixation of fractures of the distal tibia Model/Catalog Number: 770715161

DateFeb 21, 2025 (initiated Jan 21, 2025)
StatusOpen, Classified
ReasonComplaints of the locking screw passing through the locking hole intra-operatively during use of the A.L.P.S. mvX Medial Tibia plates and Anatomic Lateral Fibula. r surgical delays have been reported with the use of the devices
Root causeProcess control
ActionTyber Medical issued Urgent Medical Device Recall letter on 1/24/25 via email to the Distributor. Letter states reason for recall, health risk and action to take: 1. Acknowledge the receipt of this communication via email correspondence to feedback@tybermed.com. 2. Further distribute this communication to all applicable parties and/or customers that possess the impacted inventory. 3. Identify via laser-marked lot number and quarantine the impacted plates under Zimmer Biomet s control that have not been distributed for sale. An RMA will be provided for the return of goods to Tyber Medical for
Quantity100 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K232652Anatomical Plating System (various)
Tyber Medical, LLC, Dec 13, 2023