Recall Z-1272-2025 ยท NHA Abutment, Implant, Dental, Endosseous

Hiossen Inc.

ET Rigid Abutment (Mini) SIze: 4.5D 3.0G/H 7.0H - Intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. Model/Catalog Number: ETRGA4537MP

DateMar 5, 2025 (initiated Jan 27, 2025)
StatusOpen, Classified
ReasonPatient chart stickers provided inside the packaging may incorrectly identify ET Rigid Abutments as Regular version whereas the actual product is a Mini version; and identify the ET Multi Abutments as Mini version whereas the actual product is a Regular version
Root causeUnder Investigation by firm
ActionHoissen issued URGENT: MEDICAL DEVICE RECALL-CORRECTION letter on 1/27/25 via USPS Certified Mail. Letter states reason for recall, health risk and action to take. Immediate Actions to be taken by You, the clinician: 1. Do not use the affected devices 2. Quarantine (separate) the affected devices in a different location, away from normal stock room. 3.Immediately fill out the Acknowledgement and Receipt Form* (Fax: 267-795-7031 or Email: recall@hiossen.com ) *Response is required regardless if the devices have already been return or implanted 4. Return quarantined devices to the following addr
Quantity19
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K130662ET PROSTHETIC SYSTEM
Osstem Implant Co., Ltd., Jan 10, 2014