Recall Z-1198-2025 ยท DWJ System, Thermal Regulating

Augustine Temperature Management, LLC

Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 127 cm (50"), REF US502, to prevent or treat hypothermia

DateFeb 20, 2025 (initiated Dec 16, 2024)
StatusOpen, Classified
ReasonThere is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.
Root causeDevice Design
ActionAugustine Temperature Management began notifying its consignees on 12/16/2024 via email and visit to exchange the affected product.
Quantity1 unit
DistributionUS Nationwide distribution in the states of TX, AL, NV; AU, NZ.