Recall Z-1110-2025 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Philips North America Llc
IntelliVue Patient Monitor MX550. Model Number: 866066
| Date | Feb 5, 2025 (initiated Jan 15, 2025) |
| Status | Open, Classified |
| Reason | Potential failure to alarm for "No Device Data" INOP. The device to alarm after a Hamilton ventilator is disconnected from the EC10 Module or IntelliBridge I/O board. |
| Root cause | Software design |
| Action | Philips began notifying consignees on about 01/15/2025 via certified letters titled URGENT Field Safety Notice and URGENT Medical Device Correction letters. Consignees were instructed to ensure that there is no interruption of the cable connection between the Hamilton ventilator and the EC10 Module or IntelliBridge I/O board before use, provide the notification to all affected personnel and customers if the devices have been further distributed, and to complete and return the response form provided. A Philips representative will contact you to schedule a visit from a Philips Field Service E |
| Quantity | 74 units |
| Distribution | US distribution to Florida, North Carolina, and Oregan. International distribution to Canada, Estonia, France, Germany, Japan, Netherlands, Spain, Switzerland, United Kingdom. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K082583 | INTELLIVUE PATIENT MONITOR, INTELLIBRIDGE MODULE, MODELS MP40, MP50, MP60, MP70, MP80, MP90, EC10, EC5 Philips Medizinsysteme Boeblingen GmbH, Cardiac AN, Feb 25, 2009 |