Recall Z-1157-2025 ยท OWB Interventional Fluoroscopic X-Ray System
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Azurion 7 B12; Catalog numbers: (1) 722067, (2) 722225, (3) 722235 (OUS ONLY).
| Date | Feb 20, 2025 (initiated Jan 13, 2025) |
| Status | Open, Classified |
| Reason | Possibility of the patient falling from the table related to the mattress used on the patient table. |
| Root cause | Under Investigation by firm |
| Action | On January 14, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. 4. Actions that should be taken by the customer / user in order to prevent risks for patients or users " Circulate this Urgent Medical Device Correction letter to all users of the system so that they are aware of the issue. " Follow the additional instructions on the use of the mattress included in Appendix B. These instructions include information for the transfer of the patient and the correct use and positioning of the mattress. The information of this Appendix is being included in an Addendum to the Instru |
| Quantity | 602 units (94 US, 508 OUS) |
| Distribution | Worldiwde distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia |
| FDA record | Open on fda.gov |