Recall Z-1174-2025 ยท BYY Tube, Aspirating, Flexible, Connecting

DeRoyal Industries Inc

DeRoyal AMG Suction Connector Tubing REF: 71-2083, 71-C506, 71-C510, 71-C512, 71-C520, Tubing is intended for fluid movement

DateFeb 14, 2025 (initiated Jan 15, 2025)
StatusOpen, Classified
ReasonThe tubing may not ensure that the yankauer stays intact, and secure to the tubing / yankauer connection for a tight seal. This could result in inadequate suction and cause delay in patient care.
Root causeDevice Design
ActionOn 01/15/2025 the firm sent via email an "URGENT! DEROYAL RECALL NOTICE" letter for the Suction Connector Tubing. The letter was informing customers that the suction connector tubing could be defective and that the tubing may not ensure that the yankauer stays intact, and secure to the tubing/yankauer connection for a tight seal. Customers are instructed to: 1.Search their inventory for the affected products and place them in quarantine to prevent further use. 2. Destroy all affected products and complete the NOTICE OF RETURN FORM and indicate all affected products found in their inventory.
Quantity224,770 tubes/connectors
DistributionU.S.: ND, NY, VA, WV, NJ, TN, OK, TN, GA, OH, MN, AZ, WI, IL, MA, MD, MO, AR, MS, NY, MS, SC, HI, AR, UT, HI, O.U.S.: Kuwait
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.