Recall Z-1252-2025 ยท LLZ System, Image Processing, Radiological

Philips Medical Systems Nederland B.V.

Philips IntelliSpace Cardiovascular software, Model 830089.

DateFeb 28, 2025 (initiated Apr 4, 2024)
StatusOpen, Classified
ReasonStudy data is not able to be archived, copied, or exported with the cardiovascular software version.
Root causeSoftware design
ActionThe recalling firm initially issued an "Important Product Notice" dated 3/12/2024 which their consignees received on 4/4/2024 informing them that Philips had identified an issue with the IntelliSpace Cardiovascular software version 7.0 where study data is not able to be archived, copied, or exported. The notice was to explain the issue and under what circumstances it can occur, the affected products and how to identify them, the actions the customer can take to minimize the effect of the issue, and the actions planned by Philips to correct the issue. The actions were: (1) The customer may co
Quantity139
DistributionUS Nationwide distribution in the states of AK, AL, AZ, CA, CT, FL, GA, IA, IN, KY, MA, MI, MN, NC, NE, NY, OH, OR, TN, TX, VA, and WI. There was government distribution but no military distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K153022Philips IntelliSpace Cardiovascular
Philips Medical Systems Nethrlands BV, Nov 18, 2015