ClearanceBrief
Recall Z-1097-2025 ยท GAW Suture, Nonabsorbable, Synthetic, Polypropylene

Ethicon, LLC

PERMA-HAND Silk Suture, REF: K872H, W723H, 623H, K833H; Perma-hand Silk Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurosurgical procedures.

DateFeb 4, 2025 (initiated Dec 20, 2024)
StatusOpen, Classified
ReasonSutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine, and if issue is undetected could lead to breach in sterility, could introduce pathogens to the patient and cause infection.
Root causeProcess control
ActionOn 12/20/2024, Johnson & Johnson Med Tech mailed recall notices to Operating Room Supervisors, Materials Management Personnel, Chief of Surgery, and distributors who were asked to do the following: 1) Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). 2) Communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Pl
Quantity78792
DistributionWorldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, LA MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and the countries of AT, AU, BE, CA, CH, CN, CZ, DE, DK, FI, FR, GB, HU, IE,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K946173STAINLESS STEEL, GUT, POLY (GLYCOLIDE/L-LACTIDE) POLYAMIDE, POLYPROPYLENE, POLY (ETHYLENE TEREPHTHALATE) AND SILK
ETHICON, Inc., Jan 9, 1995