Recall Z-1113-2025 ยท PER Automated Pipetting, Diluting And Specimen Processing Workstations For Flow Cytometric Analysis

Sysmex America, Inc.

Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product usage for automated pipetting, diluting, and specimen processing workstations for flow cytometric analysis.

DateFeb 5, 2025 (initiated Jan 6, 2025)
StatusOpen, Classified
ReasonSysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.
Root causeSoftware design
ActionA Product Notification Field Correction letter was sent to customers beginning 1/6/25. Actions: 1. Until a Sysmex representative contacts you, please do not use the instrument running PS-10 software version 1.5 or 1.6 unless it is used only to create cocktails. 2. Review the content of this communication with your facility's physician and/or pathologist and retain this letter for any future reference. 3. Please fill out the form below and email it back to Sysmex using the email address: FlowPS10@Sysmex.com Questions If you have any questions concerning this information, please contact the Te
Quantity32 units
DistributionUS Nationwide.