ClearanceBrief
Recall Z-1067-2025 ยท LYJ Stimulator, Autonomic Nerve, Implanted For Epilepsy

LivaNova USA, Inc.

Brand Name: SenTiva" Product Name: VNS Therapy" SenTiva", Model 1000 Model/Catalog Number: 1000 Software Version: NA Product Description: The LivaNova VNS Therapy" System, used for Vagus Nerve Stimulation (VNS), consists of an implantable VNS Therapy generator, lead, and external programming system used to change stimulation settings. The generator is an implantable, multi-programmable pulse gener

DateJan 30, 2025 (initiated Jan 3, 2025)
StatusOpen, Classified
ReasonTheir is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.
Root causeDevice Design
ActionOn January 3, 2025, LivaNova issued a "Urgent Medical Device Removal" via Mail to affected consignees. LivaNova ask consignees to take the following actions: 1. Monitor the patient for changes in clinical symptoms or if the patient loses perception of stimulation (e.g., the typical cadence for patient visits range from every 3-12 months). 2. Perform system diagnostic testing at each patient office visit to monitor generator battery status. 3. Counsel patients to do the following: i. Notify you if there is a change in perceived clinical symptoms ii. Notify you if they no longer perceive a
Quantity8223 units
DistributionWorldwide - US Nationwide distribution including in the states of AK, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR,PA, PN, SC, SD, TN, TX, VA, VT, WA, WI, WV, UT and the countries of AE, AR, AT, BE, BG, BR, CA, CL, CO, CR, CZ, DE, DK, GR,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.