ClearanceBrief
Recall Z-1008-2025 ยท LXH Orthopedic Manual Surgical Instrument

Integrity Implants Inc.

Brand Name: N/A Product Name: Repeater Bone Tamp Model/Catalog Number: ASY-00132 Software Version: N/A Product Description: The Repeater Bone Tamp is a cylindrical shaft used to push bone graft material through a corresponding bone funnel into the FlareHawk interbody fusion device in spinal fusion procedures. Component: No

DateJan 22, 2025 (initiated Dec 20, 2024)
StatusOpen, Classified
ReasonDue to incompatibility of bone tamp (Rev A) and bone funnels (Rev C).
Root causeDevice Design
ActionOn 12/20/2024, the firm emailed a "NOTICE OF RECALL" letter to customer informing them that the firm has received a report relating to a specific lot of Repeater Bone Tamps not fitting with newer lots of the Repeater Bone Funnel. Customers are instructed to: 1. Segregate Product. immediately remove all Repeater Bone Tamp (Part # ASY-00132, Lot Number 03200045), from their inventory (regardless of its location) and segregate this product in a secure location for return to Accelus. 2. Complete QF-00070 Product Tracking & Verification Form. Complete and return the enclosed QF-00070 Product Track
Quantity9
DistributionU.S. Nationwide distribution in the states of CA, F, GA, KY, LA, MA, MD, TN and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.