Recall Z-1039-2025 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Zimmer, Inc.
NexGen LCCK Legacy Articular Surface with Locking Screw - Striped Yellow/E,F - Height 10 mm, Item Number 00-5994-032-10
| Date | Jan 28, 2025 (initiated Jan 6, 2025) |
| Status | Open, Classified |
| Reason | It was identified internally that two commingle events occurred where the metal support post within the polyethylene articular surface was assembled incorrectly for three units within the scope. |
| Root cause | Process change control |
| Action | An URGENT MEDICAL DEVICE RECALL - REMOVAL notification letter dated 1/6/25 was sent to customers. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the recall notice for hospitals and ensure documentation. 4. Complete Attachm |
| Quantity | 155 units |
| Distribution | Worldwide - US Nationwide distribution in the states of CA, FL, GA, KS, MO, NE, NJ, PA, TN and the countries of CHINA, INDIA, KOREA, MALAYSIA, NETHERLANDS, SINGAPORE, TAIWAN. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K960279 | NEXGEN COMPL. KNEE SOL. LEGACY POSTERIOR STABILIZED/CONSTRAINED CONDYLAR KNEE FEMORAL COMPONENTS/ARTICULAR SURFACES Zimmer, Inc., Apr 26, 1996 |