Recall Z-1019-2025 ยท FRN Pump, Infusion
Fresenius Kabi USA, LLC
Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.
| Date | Feb 4, 2025 (initiated Jan 10, 2025) |
| Status | Open, Classified |
| Reason | Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2. If a secondary infusion is started at the exact moment a primary infusion completes and VTBI reaches 0, it will switch to primary. The primary infusion will infuse at the previously programed primary rate and continue until the infusion is stopped or the bag is empty. |
| Root cause | Software Design Change |
| Action | Fresenius Kabi began notifying consignees on 01/10/2024 via emailed letter. Consignees were instructed to check if your device contains the affected software, review and become familiar with the temporary Risk Mitigation actions for each anomaly outlined in the letter and notify customers/personnel if the product was transferred or further distributed. Additionally, to facilitate the installation of the pump software, LVP SW 5.10, your facility must install the new Ivenix Infusion Management System (IMS) software version (5.2) on your IMS server and install the new software version (5.10) fo |
| Quantity | 23 systems |
| Distribution | US distribution to CA, CO, GA, ID, MI, MS, NE, NJ, NV, OK, TX, UT, VA, WA, WI. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K183311 | Ivenix Infusion System (IIS) Ivenix, Inc., Jun 7, 2019 |