Recall Z-1074-2025 ยท BSZ Gas-Machine, Anesthesia

Getinge Usa Sales Inc

Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i C20, Flow-i C30, Flow-i C40, Flow-c, Flow-e. For use during anesthesia.

DateFeb 7, 2025 (initiated Jan 15, 2025)
StatusOpen, Classified
ReasonSevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.
Root causeDevice Design
ActionGetinge notified consignees on 01/15/2024 via URGENT MEDICAL DEVICE RECALL letter. Consignees were provided additional information which is complementary to the existing user manual and were instructed to ensure all affected product inventory has been quarantined, consignees are entitled to credit/replacement, the recall information is forwarded to affected personnel and any customers if it was further distributed and complete and return the provided URGENT MEDICAL DEVICE RECALL RESPONSE FORM.
Quantity6 units
DistributionDomestic distribution to Kentucky and Connecticut.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K191027Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e Anesthesia System
Maquet Critical Care AB, Mar 18, 2020