Recall Z-1103-2025 ยท GEH Unit, Cryosurgical, Accessories

New Medical Technologies Gmbh

Integra Miltex CryoSolutions Cartridges 10 Pack (23.5g N2O each), Model: C-CA-23, Item: 33516;

DateFeb 5, 2025 (initiated Dec 19, 2024)
StatusOpen, Classified
ReasonDuring the assembly of the cartridge to its connecting body part, the valve may open prematurely resulting in partial or complete emptying of the cartridge before its intended use.
Root causeDevice Design
ActionOn December 19, 2024 "Important Notice Regarding Your Device" letters were sent to customers. Corrective Actions: We have implemented the following corrective and preventive actions to address this issue: 1. Enhanced Testing: All cartridges are now being filled with a minimum specified volume (23.5g tolerance) and subjected to rigorous leakage testing. 2. Updated Specifications: We have revised the interface connection specifications to include precise tolerances and correlated these adjustments with the gas volume requirements. 3. Improved Instructions for Use (IFU): Additional safety preca
Quantity430 units (43 packs of 10)
DistributionWorldwide - US Nationwide distribution in the states of AR, CA, CO, FL, GA, IL, IN, KY, LA, MA, MD, MN, MO, NC, NE, NJ, NV, NY, OH, PA, SC, TN, TX, VA, WA and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K091721CRYOSUCCESS, MODEL C-SUC-01, C-SUC-02, C-DTS-02, C-DTS-03, C-GTS-13. C-CA-23
New Medical Technologies GmbH, Oct 2, 2009