Recall Z-1111-2025 ยท NHA Abutment, Implant, Dental, Endosseous

DDS Lab

Temporary Titanium Abutments

DateFeb 5, 2025 (initiated Dec 6, 2024)
StatusOpen, Classified
ReasonCertain definitive dental implant restoration cases may have included a temporary titanium abutment (cylinder), rather than a definitive titanium abutment, which may lead the titanium abutment fracture, which may risk loosening of the affixed restoration; fragments could also injure oral mucosa, and could be aspirated.
Root causeComponent design/selection
ActionOn 12/6/2024, correction notices were mailed to customers asking them to do the following: Patients that received a dental implant restoration processed with a temporary titanium abutment (cylinder) rather than a definitive titanium abutment should be contacted to inform them about this Correction. Dental providers should continue to monitor a restoration subject to this Correction for potential problems, consistent with their clinical standard of care for all dental implant restorations. Firm is actively collaborating with dental support organizations and dental offices to ensure patients
Quantity11,259
DistributionUS Nationwide distribution including in the states of FL, TN, TX, MA, NH, CT, ME, WA, IN, CA, PA, NV, MO, IL, AK, MN, VT, OH, MI, VA, WI, AL, AZ, KY, MT, OK, CO, LA, GA, NY, WV, NC, NJ, MD, OR, DE, SC, ID, KS, IA, AR, NM, HI, DC, MS, NE, RI, UT.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071638BIOHORIZONS TAPERED INTERNAL IMPLANT SYSTEM
BioHorizons Implant Systems, Inc., Oct 10, 2007
K093321BIOHORIZONS LASER-LOK 3.0 IMPLANT SYSTEM
BioHorizons Implant Systems, Inc., Apr 2, 2010
K103691BIOHORIZONS ABUTMENTS FOR ZIMMER
BioHorizons Implant Systems, Inc., Nov 3, 2011
K223697MRI compatibility for existing BioHorizons dental implants and abutments
BioHorizons Implant Systems, Inc., Oct 13, 2023