Recall Z-0975-2025 ยท BRY Cabinet, Table And Tray, Anesthesia
CareFusion 303, Inc.
BD Pyxis Med ES Auxiliary (AUX), REF 343
| Date | Jan 17, 2025 (initiated Feb 14, 2023) |
| Status | Open, Classified |
| Reason | Drawer and/or door failures that may lead to a delay in access to medications. The following three situations in which drawer and/or door failures may occur have been identified: Situation 1: Door Latch Failure, Situation 2: Cable Ribbon Damage, Situation 3: Magnet Falling Out, |
| Root cause | Nonconforming Material/Component |
| Action | BD issued an Urgent Medical Device Product Advisory to its consignees on 02/14/2023 via letter. The notice explained the problems with the device, health hazards, and requested the following actions be taken: "1. Customers may continue to use these devices as intended and should file a complaint via the Customer Portal at https://www.bd.com/self-service or call BD Support at 1-800-727-6102, should you experience drawer or door failures. 2. If you experience delays in access to medication due to a drawer or door failure, ensure backup procedures (which may include but may not necessarily be |
| Quantity | 5918 units |
| Distribution | Worldwide |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.