ClearanceBrief
Recall Z-1125-2025 ยท BRY Cabinet, Table And Tray, Anesthesia

CareFusion 303, Inc.

BD Pyxis MedStation 4000 REF: 303. Intended to securely store and dispense medications to a qualified and authorized healthcare provider.

DateFeb 11, 2025 (initiated Feb 23, 2024)
StatusOpen, Classified
ReasonRetroactive-Due to increase in complaints, their is a potential for drawer and/or door failures of the medication cabinet which could result in delay in accessing medications.
Root causeDevice Design
ActionOn February 14, 2023 and January 9. 2024 Carefusion issued a Urgent medical Device Product Advisory" Recall Notification to affected consignees. Carefusion asked consignees take the following actions: 1. Customers may continue to use these devices as intended and should file a complaint via the Customer Portal at https://www.bd.com/self-service or call BD Support at 1-800-727-6102, should you experience drawer or door failures. 2. If you experience delays in access to medication due to a drawer or door failure, ensure backup procedures (which may include but may not necessarily be limited t
DistributionUS: Nationwide OUS:JP IT IN DE GB KR NL TW ES BE AU BH CA BR NZ FR AR QA MX CL TR SA MY CH IL IE TH SG KW MA FI AE EG BS BM GR MC PL LB MO IQ TN NO PH CN HK CO JO OM
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.