Recall Z-1005-2025 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Ortho-Clinical Diagnostics, INc.
Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: 6802413 (New), 6802415 (Certified/Refurbished), 6904001 (China) Software Version: version 3.8.0 or higher Product Description: In vitro diagnostic chemistry analyzer. FDA 3 letter product code: JJE Component: No, not a component
| Date | Jan 21, 2025 (initiated Dec 19, 2024) |
| Status | Open, Classified |
| Reason | The VITROS 5600 Integrated System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired products, which would increase the risk of no results or erroneous results from the instrument. |
| Root cause | Software Design Change |
| Action | On December 19, 2024 URGENT PRODUCT CORRECTION NOTIFICATION letters were sent to customers. REQUIRED ACTION " Do not use expired Signal Reagent packs. See the Questions & Answers section for more details. " Complete the enclosed Confirmation of Receipt form no later than December 31, 2024. " Save this notification with your User Documentation or post this notification by each VITROS 3600/5600/XT 7600 System in your laboratory until the issue has been resolved. Contact Information We apologize for the inconvenience this will cause your laboratory. If you have further questions, please contact G |
| Quantity | 3,249 units (905 US, 2,344 OUS) |
| Distribution | Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K081543 | VITROS IMMUNODIAGNOSTIC PRODUCTS TSH REAGENT PACK AND CALIBRATORS, VITROS 5600 INTEGRATED SYSTEM Ortho-Clinical Diagnostics, Inc., Oct 23, 2008 |