Recall Z-0846-2025 ยท OLO Orthopedic Stereotaxic Instrument

Tornier S.A.S.

Stryker Blueprint Mixed Reality BP MxR Glenoid Pin Guide, REF MRUE200. Indicated for use during Total Shoulder Arthroplasty.

DateJan 10, 2025 (initiated Dec 18, 2024)
StatusOpen, Classified
ReasonThe Pin Guide is being recalled because the metal tube may disassemble from the plastic handle. This can result in the Blueprint Mixed Reality headset being unable to detect the Pin Guide and potentially prolonging surgery time or changing of surgical method from guided to conventional.
Root causeDevice Design
ActionStryker issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 12/18/2024 via email. The notice explained the problem with the device and requested the following actions be taken: "1. Review the product issue as communicated in this Recall Notice & communicate the issue as appropriate in your organization. 2. Locate and isolate/quarantine the affected device to prevent accidental use pending return. 3. Please use a device from a conforming lot for all procedures. 4. Return the enclosed business reply form by email to confirm receipt of this notification. a. Response is required, eve
Quantity24 units
DistributionUS: MN, KY, MD, MA, TX, AK, TX, ID; France, Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K222510Blueprint Mixed Reality system
Tornier S.A.S., Jan 20, 2023