ClearanceBrief
Recall Z-1552-2025 ยท BRY Cabinet, Table And Tray, Anesthesia

CareFusion 303, Inc.

MEDFLEX, REF: 1119-00; MEDBANK MAIN, REF: 1137-00; MEDFLEX 2.0, REF: 1139-00; MEDBANK TOWER MAIN, REF: 1145-00; MEDBANK MINI MAIN, REF: 1147-00 BD PYXIS MEDBANK Facility Software, REF: 139090-01; BD PYXS MEDBNK TWR MN 2HH1FH2HM2FM2FG1FF, REF: 169-100; BD PYXIS MEDBANK MINI 1HH-1FH- 1HM, REF: 169-101; BD PYXIS MEDBANK MINI 1HH-1HM- 1FM, REF: 169-102; BD PYXIS MEDBANK MINI 1HH-3HM, REF: 169-103; BD

DateApr 11, 2025 (initiated Oct 5, 2023)
StatusOpen, Classified
ReasonIf "Clear All" selected medications is selected on patient profile, the automated dispensing cabinet may allow dispensing of medication without first requiring the RxVerify Validation Code be entered/without validation from a pharmacist, which could lead to unintended and incorrect medication removal.
Root causeSoftware design
Action05OCT2023, Customer Release Notes were sent to customers informing them of system release v1.3.4. On 2/18/2025, correction notices were mailed and emailed to Customers, Directors of Pharmacy, and Facility Medication Management Administrators who were informed the following: 1) Customers with system administrator privileges can uncheck the option for the "Clear All" button in the BD Pyxis MedBank Facility Software, which will disable the feature and prevent the button from displaying on the patient profile selection screen. 2) Best practice is to verify the item being removed is accurate aga
Quantity2679
DistributionUS Nationwide distribution in the states of KS, TX, CA, PA, NJ, FL, IN, AZ, IL, MA, NY, KY, UT, MD, WA, OH, AL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.