Recall Z-0831-2025 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Nuwellis Inc

nuwellis AquaFlexFlow UF 500 Plus, REF 114156, Catalog Number A06163, extracorporeal blood circuit which is used with the Aquadex SmartFlow or Aquadex FlexFlow System

DateJan 16, 2025 (initiated Dec 11, 2024)
StatusCompleted
ReasonThe AquaFlexFlow UF 500 Plus extracorporeal blood circuit used with the Aquadex SmartFlow and FlexFlow Systems may indicate "Ultrafiltrate Weight Mismatch" or Excessive Weight Mismatch Alarms while in use. If not addressed, this failure could result in excess fluid removal from a patient leading to Acute Volume Depletion. This failure is especially serious when the Aquadex System is being used on pediatric patients.
Root causeUnder Investigation by firm
ActionNuwellis issued an UREGNT: MEDICAL DEVICE RECALL notice to its consignees on 12/11/2024 via email. The notice explained the issue, risk to health, and requested the following actions be taken: Immediately discontinue use of affected lots. A Nuwellis representative will be reaching out to arrange return or disposal of the affected product to Nuwellis. For more detailed discussion regarding this product return, you may contact the Quality Team at Nuwellis: Steve Sandoval, Director of Engineering and Quality, Email: Steve.Sandoval@nuwellis.com, Office Phone: +1 952.563.7040
Quantity845 units
DistributionUS Nationwide Distribution and the country of Hong Kong
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K192756Aquadex FlexFlow System 2.0
Chf Solutions, Inc., Feb 24, 2020