Recall Z-0838-2025 ยท IKX Aid, Transfer

NOA Medical Industries Inc

NOA Medical Industries Behavioral Health Hospital bed side rails, Product number 1055050GRY-BOX

DateJan 10, 2025 (initiated Nov 5, 2024)
StatusOpen, Classified
ReasonRed button used to engage side rails can break or become stuck making it difficult to engage the side rail.
Root causeDevice Design
ActionNOA Medical Industries issued an URGENT: VOLUNTARY MEDICAL DEVICE CORRECTION notice to its consignees on 11/05/2024 via email. The notice explained the problem with the device, potential risk, and temporary actions to employ to minimize the hazard until a replacement frame is installed. A replacement side rail will be provided. For questions or concerns, contact: support@noamedical.com (email), https://www.noamedical.com/ (website) or 1-636-239-7600 (telephone), Monday-Friday 7:00AM to 3:30PM, CT.
Quantity34 units
DistributionUS: GA, MI. MA, CA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.