ClearanceBrief
Recall Z-0827-2025 ยท NKE Pulse Generator, Pacemaker, Implantable, With Cardiac Resynchronization (Crt-P)

Boston Scientific Corporation

VISIONIST CRT-P, EL (Model Number U225 and U226) and EL MRI (Model Number U228)

DateJan 10, 2025 (initiated Dec 12, 2024)
StatusOpen, Classified
ReasonA subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
Root causeProcess design
ActionThe firm sent an "URGENT MEDICAL DEVICE INFORMATION" letter dated December 2024. The letter describes the product, problem and actions to be taken. The customers were instructed to promptly identify patients, if device enters Safety Mode, perform emergent replacement for patients who are at risk for harm, during replacement of a device in Safety Mode pacing inhibition should be anticipated during electrocautery. General prophylactic replacement is not recommended; schedule device replacement promptly when the longevity remaining reaches four (4) years or if the longevity remaining is less tha
Quantity156,329 units (59,937 US, 96,792 OUS) in total
DistributionUS (Nationwide) including Puerto Rico and US Virgin Islands and OUS (foreign) to countries of: Guam and Northern Mariana Islands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.