Recall Z-0825-2025 ยท LWP Implantable Pulse Generator, Pacemaker (Non-Crt)
Boston Scientific Corporation
ALTRUA 2 DR Pacemaker, SL (Model Number S702) and EL (Model Number S722)
| Date | Jan 10, 2025 (initiated Dec 12, 2024) |
| Status | Open, Classified |
| Reason | A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance. |
| Root cause | Process design |
| Action | The firm sent an "URGENT MEDICAL DEVICE INFORMATION" letter dated December 2024. The letter describes the product, problem and actions to be taken. The customers were instructed to promptly identify patients, if device enters Safety Mode, perform emergent replacement for patients who are at risk for harm, during replacement of a device in Safety Mode pacing inhibition should be anticipated during electrocautery. General prophylactic replacement is not recommended; schedule device replacement promptly when the longevity remaining reaches four (4) years or if the longevity remaining is less tha |
| Quantity | 156,329 units (59,937 US, 96,792 OUS) in total |
| Distribution | US (Nationwide) including Puerto Rico and US Virgin Islands and OUS (foreign) to countries of: Guam and Northern Mariana Islands. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.