Recall Z-0821-2025 ยท BTL Ventilator, Emergency, Powered (Resuscitator)

Mercury Enterprises, Inc. dba Mercury Medical

Neo-Tee T-Piece Resuscitator, for pediatric emergency respiratory support, Part: 1050805, 1050808, 1050809, 1050810, 1050811, 1050814, 1050832, 1050839, 1050840, 1050841, 1050842

DateJan 10, 2025 (initiated Dec 10, 2024)
StatusOpen, Classified
ReasonGas powered emergency resuscitator has a circuit flow controller manufactured with an undersized spring, which may cause resuscitator not reach the high and low ends of the PIP and PEEP pressure ranges, which may cause a loss of positive pressure, impacting effective ventilation of the patient.
Root causeProcess change control
ActionOn 12/10/2024, "URGENT: FIELD SAFETY NOTICE" letters were mailed and emailed to medical facilities and distributors who were asked to do the following: Medical Facilities/Distributors: 1) Users should cease use and distribution of affected product and immediately quarantine the affected product. 2) Contact firm's customer service department at uscustomerservice@mercurymed.com (Domestic) and InternationalCustomerService@mercurymed.com (International) to assist you on how to send the affected product back to the firm for product replacement or credit. 3) Please notify all personnel subject to
Quantity37370
DistributionWorldwide Distribution: US (nationwide) to states of: PA, FL, IN, WI, ME, SC, WY, ID, MT, TX, MI, LA, MS, AR, WV, MD, KY, NJ, WA, CA, GA, MO, MA, NC, NV, TN, VA, OK, ND, MN, AL, HI, UT, NY, AZ, IL, OH, IA, NE, CT, AK, OR and OUS (foreign) to countries of: Hungary, Switzerland, Hong Kong, Spain, Cana
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K093913MERCURY MEDICAL T-PIECE RESUSCITATOR
Mercury Medical, Inc., Apr 6, 2010