Recall Z-0902-2025 ยท LFM Enzyme Immunoassay, Digitoxin

QUIDEL ORTHO

Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products DGXN Slides Model/Catalog Number: 8343386 Software Version: Not Applicable Product Description: The VITROS DGXN Slide is a multilayered, analytical element coated on a polyester support. Digoxin is a widely prescribed cardiac glycoside indicated in the treatment of congestive heart failure and supraventricular arrhyth

DateJan 10, 2025 (initiated Dec 5, 2024)
StatusOpen, Classified
ReasonHemoglobin interference may occur at concentrations lower than the claimed 300 mg/dL, and may cause negatively biased digoxin results.
Root causeDevice Design
ActionOn December 5, 2024 URGENT PRODUCT CORRECTION NOTIFICATION letters were sent to consignees. REQUIRED ACTIONS - For customers who have VITROS 5,1 FS/4600/5600/XT 3400/XT 7600 Systems: Load ADD DRV 6308 upon receipt. - For customers who have VITROS 250/350 Systems: Please be aware of the lowering of the Interferent Concentration limit from 300 mg/dL to 100 mg/dL at a digoxin concentration of 2.0 ng/mL (2.6 nmol/L). - Complete and return the enclosed Confirmation of Receipt form no later than December 13, 2024. - Save this notification with your User Documentation or post this notification by eac
Quantity1,541 units (1,125 US, 416 OUS)
DistributionDomestic: Nationwide Distribution International: Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom,
FDA recordOpen on fda.gov